K-6
Operator’s Manual – OPTI CCA-TS2
POTASSIUM
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
54
Range
0.9 to 8.1
Slope
1.01
Offset
-0.01
R
2
0.999
Sy|x
0.031
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Tsien R,
New Calcium Indicators and Buffers with High Selectivity Against Magnesium and Protons
,
Biochemistry 19, p.2396-2404, 1980
.
3. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
K
-
TS2
K - TS
Scatter Plot with Fit
Identity
Linear fit
(-0.01 + 1.01x)
Содержание CCA-TS2