Operator’s Manual – OPTI CCA-TS2
Lac-5
ANALYTES
LACTATE
Parameter and Specification
Value
Number of Samples
55
Slope
1.04
Offset
-0.02
R
2
0.995
Sy|x
0.390
Correlation to Other Methods
2
OPTI CCA-TS vs other Lactate Instruments on whole blood in a typical setting
Excess sample aliquots from specimens collected for metabolyte analyses were analyzed by both
traditional and non-traditional operators of measurement equipment in hospital laboratories. The sample
was analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation
used for these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood)
0.96
-0.01
0.944
0.64
0.78 to 13.96 175
Analyzer B (whole blood)
1.18
-0.54
0.953
0.55
0.40 to 11.3
49
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
0
2
4
6
8
10
12
14
16
18
20
0
2
4
6
8
10
12
14
16
18
20
La
c
-T
S2
Lac - TS
Scatter Plot with Fit
Identity
Linear fit
(-0.02 + 1.04x)