Operator’s Manual – OPTI CCA-TS2
A-1
APPENDIX A - TECHNICAL SPECIFICATIONS
APPENDIX A - TECHNICAL SPECIFICATIONS
Measurement Range
Parameter
Range
Display
Units
Resolution
(Lo/Hi)
pH
6.6 to 7.8
0.01/0.001 pH units
P
CO
2
10 to 200
1/0.1
mmHg
P
O
2
10 to 700
1/0.1
mmHg
Na
+
100 to 180
1/0.1
mmol/L
K
+
0.8 to 10
0.1/0.01
mmol/L
Ca
++
0.2 to 3.0
0.01
mmol/L
Cl-
50 to 160
1/0.1
mmol/L
Glu
30 to 400
0.1
mg/dL
(70 to 400 for samples with
P
O
2
levels between 401-700 mmHg)
Glu
1.7 to 22
0.01
mmol/L
BUN
2.8 to 112
0.1
mg/dL
Urea
1 to 40
0.01
mmol/L
Lac
0.3 to 17.5
0.01
mmol/L
tHb
5 to 25
0.1
g/dL
SO
2
60 to 100
1/0.1
%
Barometric Pressure
300 to 800 mmHg
Operating Altitude
Up to 3048m (10,000ft)
Pollution Degree
Degree 2, normal indoor laboratory environment. Air contains only non-conductive
pollutants with occasional condensation.