Operator’s Manual – OPTI CCA-TS2
3-3
3 CUSTOMIZATION
3.2.1 QC Setup
3.2.1.1 Setting up the Quality Control Material
Fig. 3-6 Select Control
When you open a new box of OPTI CHECK or
OPTI CHECK
PLUS
, the lot number should be
entered into the analyzer, along with the target
ranges. Each QC level of control has its own
unique lot number printed on the information sheet
contained in the control box.
NOTE: OPTI CHECK and OPTI CHECK PLUS
Quality Control materials are designed for
your OPTI CCA-TS2 and have assigned
assay ranges for each measured parameter
.
NOTE: The procedure for programming QC ranges
as described below is identical for all levels.
NOTE: The OPTI CCA-TS2 can save information
for one lot of OPTI CHECK and one lot of
OPTI CHECK PLUS concurrently for each
level.
1. In the main menu, select
<System Manager>
and
<Setup>
.
2.
Enter security information if enabled
(see Section 3.2.3.1).
3. On the
<Patient/QC>
tab, select
<Control>
(Fig. 3-6).
4. Take the bar code sheet out of the OPTI CHECK
box and scan
Barcode A
for the applicable
level of OPTI CHECK or OPTI CHECK
PLUS
(Fig. 3-7).
•
Hold the bar code 2-3 inches (5-8 cm) from
the bar code scanner located on the bottom
right-hand corner of the analyzer.
•
The red line from the bar code scanner
should cover the entire bar code.
•
A beep and a green status light indicates a
valid bar code.
Fig. 3-7 Scan Bar Code