Operator’s Manual – OPTI CCA-TS2
8-33
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.21 Cassette Detect
The purpose of this test is to calibrate the Cassette
Detect function.
In the main menu, press
<System Manager ->
Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Cassette Detect>
to start the
calibration (Fig. 8-41).
•
Insert a patient cassette (Fig. 8-42).
•
Remove the cassette and leave the cover open
(Fig. 8-43).
•
Close the SMC cover (Fig. 8-44).
•
The system performs a Cassette Detection
Calibration (Fig. 8-45).
Fig. 8-41 Cassette Detect
Fig. 8-42 Insert Cassette
Fig. 8-44 Close Cover
Fig. 8-45 Calibration
Fig. 8-43 Remove Cassette
Содержание CCA-TS2