Cl-4
Operator’s Manual – OPTI CCA-TS2
CHLORIDE
ANALYTES
Direct linearity for the OPTI CCA-TS2 measurement has been established against reference materials
or methods. The linearity of the Chloride sensor on the OPTI CCA-TS2 has been established versus the
same sensor on the OPTI CCA-TS using whole blood samples tonometered with mid level mixtures and
spiked or diluted to establish the correlation.
Parameter and Specification
Value
Number of Samples
63
Slope
0.99
Offset
-0.21
R
2
0.999
Sy|x
1.218
40
60
80
100
120
140
160
180
40
60
80
100
120
140
160
180
C
l -
TS2
Cl - TS
Scatter Plot with Fit
Identity
Linear fit
(-0.21 + 0.99x)