8-28
Operator’s Manual – OPTI CCA-TS2
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.17 Checking the Display
The purpose of this test is to check the proper
operation of the display.
In the main menu, press
<System Manager ->
Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Display>
to start the test (Fig. 8-29).
•
The display will turn red, green and blue.
If this is not the case, your display is defective
and needs to be replaced.
•
Press
to return to the
<System>
screen
or
to return to the main menu.
Fig. 8-29 Display Test
8.2.16 Checking the Pump Motor
The purpose of this test is to check the proper
functioning of the peristaltic pump motor.
In the main menu, press
<System Manager ->
Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Pump>
to start the test (Fig. 8-27).
•
The pump will automatically step through all the
speeds used during normal operation (7.5 to 120
rpm (revolutions per minute)) (Fig. 8-28) and
return to the
<Tests>
screen.
Fig. 8-27 Pump Motor Test
Fig. 8-28 Pump Speed