Operator’s Manual – OPTI CCA-TS2
8-27
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.15 Checking the Pump Flow
This option is designed to test the pump cartridge.
In the main menu, press
<System Manager ->
Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Flow>
to start the test (Fig. 8-22).
•
Insert a new cassette (Fig 8-23).
•
Close the SMC cover.
•
Wait for test results (Fig. 8-24).
•
The two numbers indicate the actual flow rates
clockwise and counter clockwise (Fig. 8-25).
If one of the two or both rates are out of range,
the test fails.
•
Repeat test or replace the pump cartridge, if
the test fails. See replacement instructions in
Chapter 7.4.1.
NOTE: It is possible that the test fails the first
time, even if the pump cartridge is working
correctly.
•
Remove the cassette (Fig. 8-26).
Fig. 8-22 Flow Test
Fig. 8-24 Flow Test in Process
Fig. 8-23 Insert Cassette
Fig. 8-25 Flow Test Pass
Fig. 8-26 Remove Cassette