
PO2-2
Operator’s Manual – OPTI CCA-TS2
P
O
2
ANALYTES
Reproducibility
Controls
Within run precision (S
wr
) and Total Precision (S
T
), were determined from 2 runs per day with 2 replicates
per run over a period of 20 days following the CLSI guideline EP5-A2. Typical results for 3 different
control levels are shown below.
PO
2
(mmHg)
OPTI Check Level 1
OPTI Check Level 2
OPTI Check Level 3
Days Run
20
20
20
Total Average
71.6
100.0
137.7
Within Run St. Dev. (S
wr
)
1.3
1.4
1.4
Within Run % CV
1.8%
1.4%
1.0%
Total Precision St. Dev. (S
T
)
1.5
1.7
1.8
Total % CV
2.1%
1.7%
1.3%
Whole Blood
Within-run precision in whole blood samples was evaluated at 3 different concentrations on the OPTI
CCA-TS2. The table below shows the Averages, Standard Deviations and % CV calculated from 10
samples for each level.
PO
2
(mmHg)
Whole Blood
Level 1
Level 2
Level 3
Average
54.8
90.3
434.5
St. Dev
0.5
0.5
5.1
%CV
0.97%
0.52%
1.18%
n
10
10
10
Содержание CCA-TS2
Страница 1: ...OPTI CCA TS2 Analyzer Operator s Manual ...