Operator’s Manual – OPTI CCA-TS2
PCO2-B-3
ANALYTES
P
CO
2
(Dry Sensor)
Reproducibility
Controls
Within run precision (S
wr
) and Total Precision (S
T
), were determined from 2 runs per day with 2 replicates
per run over a period of 20 days following the CLSI guideline EP5-A2. Typical results for 3 different
control levels are shown below:
Dry PCO
2
(mmHg)
OPTI Check Level 1
OPTI Check Level 2
OPTI Check Level 3
Days Run
20
20
20
Total Average
72.6
43.1
22.5
Within Run St. Dev. (S
wr
)
0.8
0.3
0.3
Within Run % CV
1.1%
0.7%
1.4%
Total Precision St. Dev. (S
T
)
1.2
0.6
0.4
Total % CV
1.6%
1.3%
2.0%
Whole Blood
Within-run precision in whole blood samples was evaluated at 3 different concentrations on the OPTI
CCA-TS2. The table below shows the Averages, Standard Deviations and % CV calculated from 10
samples for each level.
Dry PCO
2
(mmHg)
Whole Blood
Level 1
Level 2
Level 3
Average
19.6
44.4
92.8
St. Dev
0.357
0.316
1.088
%CV
1.82%
0.71%
1.17%
n
10
10
10