pH-B-4
Operator’s Manual – OPTI CCA-TS2
pH (Dry Sensor)
ANALYTES
Linearity
Wherever possible, linearity for the OPTI CCA-TS2 measurement has been established against reference
materials or methods. The linearity of the pH sensor on the OPTI CCA-TS2 has been established versus
the same sensor on the OPTI CCA-TS using whole blood samples tonometered with different %CO
2
gas
mixtures to establish the correlation.
Parameter and Specification
Value
Number of Samples
81
Slope
0.99
Offset
0.09
R
2
0.998
Sy|x
0.015
6.4
6.6
6.8
7
7.2
7.4
7.6
7.8
8
6.4
6.6
6.8
7
7.2
7.4
7.6
7.8
8
D
ry
p
H
-
TS2
Dry pH - TS
Scatter Plot with Fit
Identity
Linear fit
(0.09 + 0.99x)