4-8
Operator’s Manual – OPTI CCA-TS2
4 CALIBRATION AND QUALITY CONTROL
4.5.2 Running a QC Sample
Policies regarding the measurement of QC samples
are at the discretion of the individual hospital. OPTI
Medical Systems recommends that QC solutions
be run, as a minimum, with each new lot number of
cassettes and at monthly intervals thereafter.
You should only use the manufacturer recommended
controls OPTI CHECK and OPTI CHECK P
LUS
which do
NOT
contain dye or other colored material.
Whenever a new lot of controls is opened, be sure to
enter the lot number information into the analyzer as
described in Chapter 3 “Customization”.
NOTE: Store controls at temperature recommended
by the manufacturer
NOTE: The target value of PO
2
is very sensitive to
storage conditions and barometric pressure.
High altitude environments may see recovery
outside the target range.
The control material should provide target values
for all measured parameters over a range of
measurement values typically seen in a laboratory.
The results obtained should fall within limits
established by the user’s laboratory.
4.5.2.1 Running Controls (OPTI CHECK, OPTI CHECK PLUS)
1. To run controls, press
<QC Manager>
in the
main menu (Fig. 4-12), and select
<Control>
in the
<QC>
menu.
Fig. 4-12 Select QC-Manager