A-2
Operator’s Manual – OPTI CCA-TS2
APPENDIX A - TECHNICAL SPECIFICATIONS
Operating Parameters
Minimum Sample Size
125µL (60µL for B60 cassette)
Sample Type
heparinized whole blood, plasma or serum
Sample Application
syringe, capillary or ComfortSampler
Sample Input
automatic aspiration
Analysis Time
< 2 minutes, typically approx. 1 minute
to result
Ambient Temperature Range
10 °C - 30 °C (50 °F - 86 °F)
Relative Humidity Range
5% - 95% (non-condensing)
Type of Measurement
optical fluorescence,
for tHb/SO
2
optical absorbance/reflectance
Input Values
Patient ID
25 alphanumeric characters
Accession Number
25 alphanumeric characters
Date of birth
Month, DD, YYYY
Patient Sex
Male, female or unknown
Patient Temperature
14.0 – 44.0° C (57.2 - 111.2°F)
Medical Record Number
25 alphanumeric characters
Account Number
25 alphanumeric characters
Test ID
25 alphanumeric characters
Patient Name
25 alphanumeric characters
Age
1-150
Attending Physician
25 alphanumeric characters
Patient Location
25 alphanumeric characters
Sample Collection Time
Month, DD, YYYY, HH:MM
Sample Type
Art/Ven/MixVen/Cap/Cord/CPB, where:
Art = Arterial
Ven = Venous
MixVen = Mixed Venous
Cap = Capillary
Cord = Cord
CPB = Cardio-Pulmonary Bypass
Puncture Site
LR/RR/LB/RB/LF/RF/Cord/Scalp, where:
LR/RR = Left Radial/Right Radial
LB/RB = Left Brachial/Right Brachial
LF/RF = Left Femoral/Right Femoral
Cord = Cord
Scalp = Scalp
Allen’s Test
Unknown, positive, negative
Содержание CCA-TS2