Operator’s Manual – OPTI CCA-TS2
7 MAINTENANCE
7. When this display appears (Fig. 7-25), press
to install a new gas bottle.
NOTE: If you are reinstalling a used bottle, respond
to the
<New Gas Bottle?>
prompt. You will then be asked to enter
the number of weeks in service using the
numeric keypad (Fig. 7-26). Here you may
refer back to the installation date, which was
recorded on the gas bottle.
NOTE: The gas bottle in-use expiration is 6 months
from installation or the shelf life of the gas
bottle, whichever comes first.
•
The analyzer will initiate a purge of the system,
which will be indicated by a progress bar
displayed on the screen, and will then return to
the main screen (Fig. 7-27).
Fig. 7-25 New Gas Bottle
Fig. 7-26 Number of Weeks in use
Fig. 7-27 Ready screen