9-6
Operator’s Manual – OPTI CCA-TS2
9 OPERATING PRINCIPLES
9.5.2 Test Conditions
Sample Size:
a minimum of 125 µL (60µL for B60 cassette)
Sample Type:
heparinized whole blood, serum and plasma
Sample Application:
syringe, capillary or ComfortSampler
Ambient Temperature:
10 - 30 ºC (50 – 86 ºF)
Relative Humidity:
5% to 95% (non-condensing)
Type of Measurement:
optical fluorescence (pH,
P
O
2
,
P
CO
2
, Na
+
, K
+
, Ca
++
) Cl
-
,
Glu, BUN (urea), Lac) and reflectance (tHb,
S
O
2
)
9.5.3 Input Values
Parameter
Ranges/Options
Default
Patient ID
25 alphanumeric characters
Blank
Accession Number
25 alphanumeric characters
Blank
Date of Birth
Month, DD, YYYY
Patient Sex
Male, female or unknown
unknown
Patient temperature, T
14.0 to 44.0 °C
37.0 °C
57.2 to 111.2 °F
98.6 °F
Medical Record Number
25 alphanumeric characters
Blank
Account Number
25 alphanumeric characters
Blank
Test ID
25 alphanumeric characters
Blank
Patient Name
25 alphanumeric characters
Blank
Age
1-150
0
Attending Physician
25 alphanumeric characters
Blank
Patient Location
25 alphanumeric characters
Blank
Sample Collection Time
Month, DD, YY, HH:MM
Blank
Sample Type
Art, Ven, MixVen, Cap,
Art
Cord, CPB, where:
Art = Arterial
Ven = Venous
MixVen = Mixed Venous
Cap = Capillary
Cord = Cord
CPB = Cardio-Pulmonary Bypass
Содержание CCA-TS2