Operator’s Manual – OPTI CCA-TS2
Glu-5
ANALYTES
GLUCOSE
Correlation to Other Methods
2
OPTI CCA-TS vs other Glucose Instruments on Whole Blood, Plasma and Serum in a
typical setting
Excess sample aliquots from specimens collected for metabolyte analyses were analyzed by both
traditional and non-traditional operators of measurement equipment in hospital laboratories. The sample
was analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation
used for these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood) mg/dL
1.0079
-0.7539
0.9932
6.509
30 to 400
138
Analyzer B (plasma) mg/dL
0.9986
-2.34
0.9866
8.500
44 to 398
167
OPTI CCA mg/dL
(whole blood vs. plasma) mmol/L
1.058
1.058
2.36
0.13
0.9700
0.9700
21.600
1.200
37 to 395
2.1 to 21.9
103
103
Analyzer C (serum) mg/dL
0.9500
5.73
0.9784
10.51
78 to 294
68
Analyzer D (serum) mg/dL
0.9910
3.99
0.9772
10.74
36 to 344
102
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.