Operator’s Manual – OPTI CCA-TS2
PCO2-B-5
ANALYTES
P
CO
2
(Dry Sensor)
Correlation to Other Methods
2
OPTI CCA-TS vs other Blood Gas Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of blood gas equipment in hospital laboratories. The blood was analyzed
on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for these
analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
OPTI R
0.99
4.15
0.984
3.53
21 to 184
146
Analyzer A
0.94
1.88
0.982
1.62
22.7 to 93.2
112
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
OPTI CCA-TS with B-Lac Cassette vs other Blood Gas Instruments on whole blood
(in-house testing)
Whole blood samples from multiple donors were tonometered with different %CO
2
gas mixtures to
generate a wide range of PCO
2
values. The blood samples were analyzed in parallel on the B-Lac cassette
and other laboratory instruments.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
OPTI CCA
1.00
2.12
0.994
3.52
13 to 196
162
Analyzer B
0.96
1.75
0.986
3.34
13 to 104
153
Analyzer C
1.01
-0.47
0.994
3.74
15 to 199
162
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.