Glu-6
Operator’s Manual – OPTI CCA-TS2
GLUCOSE
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
54
Range
69.4 to 361.3
Slope
0.99
Offset
-0.44
R
2
0.983
Sy|x
12.849
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2.
OPTI Medical.
Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
50
100
150
200
250
300
350
400
50
100
150
200
250
300
350
400
G
lu
-
TS2
Glu - TS
Scatter Plot with Fit
Identity
Linear fit
(-0.44 + 0.99x)