8-20
Operator’s Manual – OPTI CCA-TS2
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.4 Checking the LEDs
This menu displays the current barometric pressure.
In the main menu, press
<System Manager ->
Diagnostics>
.
•
The
<Miscellaneous>
section will show the
current barometric pressure (Fig. 8-6).
•
If the barometric pressure requires adjustment,
refer to Setup, Section 3.2.4.4.1 “Entering the
Barometric Pressure” for setting the barometer.
•
Press
to return to the
<System>
screen
or
to return to the main menu.
Fig. 8-6 Barometric Pressure
8.2.5 Verifying Barometric Pressure
Fig. 8-5 LEDs
This menu can be used to check proper functioning
of the LEDs and is designed for use by trained
service personnel only.
In the main menu, press
<System Manager ->
Diagnostics>
.
The following information is displayed in the
<LEDs>
section (Fig. 8-5):
•
<Front>
,
<Rear>
,
<Ion>
- fluid light gates.
•
<Cassette Misseat>
detector (located in
cover)
•
<Cassette Detect>
sensor
•
<SMC Cover>
- this function indicates whether
the SMC cover is closed or open.
•
Press
to return to the
<System>
screen
or
to return to the main menu.
Содержание CCA-TS2