PO2-4
Operator’s Manual – OPTI CCA-TS2
P
O
2
ANALYTES
Correlation to Other Methods
4
OPTI CCA-TS vs other Blood Gas Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of blood gas equipment in hospital laboratories. The blood was analyzed
on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for these
analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood)
0.9419
3.28
0.9976
18.27
36 to 563
103
Analyzer B (whole blood)
1.0192
-4.13
0.9969
4.10
34 to 291
173
Analyzer C (whole blood) 0.918 ± 0.087 8.083 ± 1.402
0.9954
8.032
29 to 407
105
Analyzer D (whole blood) 1.041 ± 0.006 -6.244 ± 0.931
0.9969
6.379
37 to 598
174
Analyzer E (whole blood) 0.993 ± 0.009 1.646 ± 0.893
0.9925
4.458
34 to 322
183
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
Содержание CCA-TS2