Na-6
Operator’s Manual – OPTI CCA-TS2
SODIUM
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
45
Range
109.0 to 179.4
Slope
0.99
Offset
1.57
R
2
0.999
Sy|x
0.390
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Tsien R,
New Calcium Indicators and Buffers with High Selectivity Against Magnesium and Protons
,
Biochemistry 19, p.2396-2404, 1980
.
3. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).