Operator’s Manual – OPTI CCA-TS2
THB/SO2-7
ANALYTES
tHb/SO
2
Correlation to Other Methods
4
OPTI CCA-TS vs other pH/Blood Gas Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of blood gas equipment in hospital laboratories. The blood was analyzed
on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for these
analyses and operated and controlled following their established procedures.
Correlation tHb
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood)
1.0284
-0.375
0.9778
0.47
6.0 to 16.1
103
Analyzer B (whole blood)
0.9877
0.14
0.9715
0.37
6.9 to 14.8
173
Analyzer C (whole blood)
1.077 ± 0.020 -0.284 ± 0.227
0.9650
0.74
5.4 to 17.4
215
Correlation SO
2
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood)
0.8678
12.99
0.9738
0.73
73 to 100
103
Analyzer B (whole blood)
0.7972
18.80
0.9064
1.81
64 to 100
173
Analyzer C (whole blood)
1.021 ± 0.016 -2.920 ± 1.522
0.9752
1.47
62 to 100
215
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.