PCO2-B-4
Operator’s Manual – OPTI CCA-TS2
P
CO
2
(Dry Sensor)
ANALYTES
Linearity
Wherever possible, linearity for the OPTI CCA-TS2 measurement has been established against reference
materials or methods. The linearity of the pCO
2
sensor on the OPTI CCA-TS2 has been established versus
the same sensor on the OPTI CCA-TS using whole blood samples tonometered with different %CO
2
gas
mixtures to establish the correlation.
Parameter and Specification
Value
Number of Samples
81
Slope
1.01
Offset
-1.19
R
2
0.998
Sy|x
1.973
0
50
100
150
200
250
0
50
100
150
200
250
D
ry
PC
O
2
-T
S2
Dry PCO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(-1.19 + 1.01x)