PO2-B-4
Operator’s Manual – OPTI CCA-TS2
P
O
2
(Dry Sensor)
ANALYTES
Linearity
Wherever possible, linearity for the OPTI CCA-TS2 measurement has been established against reference
materials or methods. The linearity of the pO
2
sensor on the OPTI CCA-TS2 has been established versus
the same sensor on the OPTI CCA-TS using whole blood samples tonometered with different %O
2
gas
mixtures to establish the correlation.
Parameter and Specification
Value
Number of Samples
72
Slope
0.99
Offset
0.43
R
2
0.999
Sy|x
3.528
0
100
200
300
400
500
600
700
0
100
200
300
400
500
600
700
D
ry
PO
2
-T
S2
Dry PO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(0.43 + 0.99x)