Operator’s Manual – OPTI CCA-TS2
K-5
ANALYTES
POTASSIUM
Correlation to Other Methods
3
OPTI CCA-TS vs other Electrolyte Instruments on Whole Blood and Serum in a typical
setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of measurement equipment in hospital laboratories. The sample was
analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for
these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood)
1.0816
-0.138
0.9857
0.13
2.1 to 6.4
103
Analyzer B (whole blood)
1.0225
-0.008
0.9673
0.15
2.4 to 6.0
173
Analyzer C (whole blood) 1.021 ± 0.019
-0.087 ± 0.077
0.9830
0.197
2.3 to 9.4
105
Analyzer D (whole blood) 1.050 ± 0.126
0.062 ± 0.055
0.9879
0.055
2.2 to 9.4
174
Analyzer E (serum)
1.084
-0.315
0.9855
0.181
2.9 to 7.5
68
Analyzer F (serum)
1.126
-0.397
0.9784
0.108
3.0 to 5.4
102
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.