Ca-4
Operator’s Manual – OPTI CCA-TS2
CALCIUM
ANALYTES
Direct linearity for the OPTI CCA-TS2 measurement has been established against reference materials or
methods. The linearity of the Ionized Calcium sensor on the OPTI CCA-TS2 has been established versus
the same sensor on the OPTI CCA-TS using whole blood samples tonometered with mid level mixtures
and spiked or diluted to establish the correlation.
Parameter and Specification
Value
Number of Samples
72
Slope
0.99
Offset
0.01
R
2
0.999
Sy|x
0.017
0
0.5
1
1.5
2
2.5
3
3.5
0
0.5
1
1.5
2
2.5
3
3.5
C
a
-T
S2
Ca - TS
Scatter Plot with Fit
Identity
Linear fit
(0.01 + 0.99x)