Operator’s Manual – OPTI CCA-TS2
PO2-B-3
ANALYTES
P
O
2
(Dry Sensor)
Reproducibility
Controls
Within run precision (S
wr
) and Total Precision (S
T
), were determined from 2 runs per day with 2 replicates
per run over a period of 20 days following the CLSI guideline EP5-A2. Typical results for 3 different
control levels are shown below.
Dry PO
2
(mmHg)
OPTI Check Level 1
OPTI Check Level 2
OPTI Check Level 3
Days Run
20
20
20
Total Average
73.5
103.5
139.1
Within Run St. Dev. (S
wr
)
1.4
0.9
1.4
Within Run % CV
1.8%
0.9%
1.0%
Total Precision St. Dev. (S
T
)
1.6
1.3
2.0
Total % CV
2.2%
1.3%
1.5%
Whole Blood
Within-run precision in whole blood samples was evaluated at 3 different concentrations on the OPTI
CCA-TS2. The table below shows the Averages, Standard Deviations and % CV calculated from 10
samples for each level.
Dry PO
2
(mmHg)
Whole Blood
Level 1
Level 2
Level 3
Average
55.9
91.8
428.7
St. Dev
0.6
0.5
5.8
%CV
1.02%
0.57%
1.36%
n
10
10
10