Operator’s Manual – OPTI CCA-TS2
PO2-3
ANALYTES
P
O
2
Linearity
Wherever possible, linearity for the OPTI CCA-TS2 measurement has been established against reference
materials or methods. The linearity of the pO
2
sensor on the OPTI CCA-TS2 has been established versus
the same sensor on the OPTI CCA-TS using whole blood samples tonometered with different %O
2
gas
mixtures to establish the correlation.
Parameter and Specification
Value
Number of Samples
90
Slope
0.99
Offset
-2.57
R
2
0.999
Sy|x
4.938
0
100
200
300
400
500
600
700
800
0
100
200
300
400
500
600
700
800
PO
2
-T
S2
PO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(-2.57 + 0.99x)