Operator’s Manual – OPTI CCA-TS2
POTASSIUM
ANALYTES
Direct linearity for the OPTI CCA-TS2 measurement has been established against reference materials or
methods. The linearity of the Potassium sensor on the OPTI CCA-TS2 has been established versus the
same sensor on the OPTI CCA-TS using whole blood samples tonometered with mid level mixtures and
spiked or diluted to establish the correlation.
Parameter and Specification
Value
Number of Samples
72
Slope
1.00
Offset
0.00
R
2
0.999
Sy|x
0.32
0
2
4
6
8
10
12
0
2
4
6
8
10
12
K
-
TS2
K - TS
Scatter Plot with Fit
Identity
Linear fit
(0.00 + 1.00x)