Operator’s Manual – OPTI CCA-TS2
3-21
3 CUSTOMIZATION
Fig. 3-47 Security Settings
The
<Security Settings>
menu has various
security options.
1. To disable all security options, select
<None>
in the
<User ID Entry>
drop-down box (Fig.
3-47). With all security disabled, the only menus
that require user ID and password entry are the
Fset (Factory Settings), Time/Date and Security
Setup menus. Time/Date and Security Setup
menus use the defaults
ADMIN/ADMIN
.
FSet requires a different User ID and password.
2.
The default security setting is
<Non-secure
User ID Entry>
(Fig. 3-48). In this mode, the
analyzer will request a user ID before patient
testing, QC testing, maintenance activities
and running an Hb calibrator. This mode
will not verify a user ID and does not require
programming of user IDs or groups. In this
mode, some activities will still be protected by
the
ADMIN
password, such as changing the time
and date and the security setup menu.
3.
<Secure User ID Entry>
(Fig. 3-49) will
enable the option to set up secure user IDs and
create groups of users with certain privileges.
This option must be selected to enable
passwords and user ID expiration.
When this option is enabled, users will be asked
to log in to the OPTI before they can perform
any operations. The OPTI will verify the login
and only allow the user to perform duties
assigned to that user. Logins can be by user ID
only.
The user ID can be entered by bar code scanner
for easy access. Once logged in, the OPTI will
enter the user ID for all activities performed by
the user automatically.
•
For added security, you can enable
passwords by selecting
<Password
Enable>
. When users first log in to the
OPTI, they will be asked to enter a password
(system default
PASSWORD
). For future
login, they will be required to enter user ID
and password.
3.2.3.1.1 Security Settings
Fig. 3-48 Non-secure User ID
Fig. 3-49 Secure User ID