8 DIAGNOSTICS AND TROUBLESHOOTING
<Not Ready: Battery Critical>
The battery is discharged and the instrument will not
perform any measurements.
•
Install a freshly charged battery or recharge for
up to 2.5 hours before the next sample is run,
or
operate the analyzer on AC power.
<Not Ready: Temperature Out Of Range>
The temperature is out of range.
•
Wait for the analyzer to reach the correct
temperature.
•
Cycle the power if the analyzer does not go to
<Ready>
within a few minutes.
•
If the analyzer does not become
<Ready>
within a reasonable time, check the temperature
under
<System - Diagnostics>
.
Check that ambient temperature is within
operating specifications on page 2-1.
<Not Ready: Temperature Error>
The SMC temperature exceeds 39 °C for more than
20 seconds.
•
Press
<Data Manager>
or
<System
Manager>
to exit this screen.
•
Contact Technical Support for assistance.
<Not Ready: SRC Lockout>
If
<SRC Lockout>
has been activated in
<Setup>
(see Section 3.2.1.2), this message will be displayed
if SRCs have not been run within the specified time.
•
Press
<Run SRC>
and run SRCs
(see Section 4.5.1).