THB/SO2-6
Operator’s Manual – OPTI CCA-TS2
tHb/SO
2
ANALYTES
The linearity of the SO
2
% measurement on the OPTI CCA-TS2 has been established versus the same
sensor on the OPTI CCA-TS using whole blood samples tonometered with different %O
2
gas mixtures to
establish the correlation.
Parameter and Specification
Value
Number of Samples
72
Slope
0.96
Offset
3.23
R
2
0.991
Sy|x
1.325
55
60
65
70
75
80
85
90
95
100
55
60
65
70
75
80
85
90
95
100
SO
2
-T
S2
SO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(3.23 + 0.96x)