PO2-B-6
Operator’s Manual – OPTI CCA-TS2
P
O
2
(Dry Sensor)
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
66
Range
10.7 to 656.6
Slope
0.99
Offset
0.36
R
2
0.999
Sy|x
3.680
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Kautsky H,
Quenching of Luminescence by Oxygen
, Transactions Faraday Society 35, p.216, 1939
3. CDI,
3M Healthcare System 200 Extracorporeal Blood Gas Monitor
. See, for example, Lubbers DW,
Gehrich J, Opitz N,
Fiber Optics Coupled Flourescence Sensors for Continuous Monitoring of Blood
Gases in the Extracorporeal Circuit,
Life Supports Systems 4, p.94, 1986.
4. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
0
100
200
300
400
500
600
700
0
100
200
300
400
500
600
700
D
ry
PO
2
-T
S2
Dry PO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(0.36 + 0.99x)