THB/SO2-8
Operator’s Manual – OPTI CCA-TS2
tHb/SO
2
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
tHb
Parameter and Specification
Value
Number of Samples
45
Range
7.7 to 21.2
Slope
1.01
Offset
-0.09
R
2
0.996
Sy|x
0.298
6
8
10
12
14
16
18
20
22
6
8
10
12
14
16
18
20
22
tH
b
-T
S2
tHb - TS
Scatter Plot with Fit
Identity
Linear fit
(-0.09 + 1.01x)