How to Use Reserved Racks
Processing Samples Manually
UniCel DxC Systems Instructions For Use A13914
Preparing Samples for Analysis
October 2005
Page 3-13
3
4
Identify samples using the table below:
If…
Then…
Sample has a readable bar
code,
• Type in the sample ID.
• If previous programming appears,
it may be necessary to clear it.
• Go to Step 2, or Step 5 if no
previous program appeared.
Sample has no bar code or bar
code can not be read,
• Type in the sample ID.
• Select the Rack field.
• Type in a rack number that is not
programmed.
• Go to Step 5.
5
If the sample is stat, select the STAT check box.
6
Select Sample Type and Sample Comment if provided.
7
To enter Demographics, select Demog [F2]. Type in the information
provided and select Next [F10] to return to the Program Sample screen.
8
Select each chemistry and/or select a panel. To access additional
chemistries, use the up and down arrows to move to other pages.
9
Select Next [F10].
10
To program additional samples, repeat Steps 2-8.
11
Get the rack(s) listed on the screen and place the samples into assigned
rack positions as shown on the load list.
12
For a Routine Sample or if system is in Standby: Priority samples
should be placed on the autoloader first.
• Place the rack in the autoloader with the rack bar code label to the right
and,
• Press RUN.
For a STAT Sample: If the system is running and there are other racks on
the autoloader,
• Press PRIORITY. The rack pusher moves back one space so the STAT
rack can be placed in front of the other racks.
• Press RUN.
Step
Action
, continued
(2 of 2)