Weekly Maintenance
Clean Probes (Exterior)
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-12
October 2005
Clean Probes (Exterior)
Follow the steps below to clean the exterior of the sample and reagent probes. Periodic
wiping of the probes is needed to clean the entire (upper and lower) probe area.
9
Calibrate Total Protein and Albumin chemistries and run QC materials.
NOTICE
A lamp calibration should be performed after Total Protein and Albumin
cup maintenance on a
MONTHLY
basis. You are not required to perform
the ALBm and TPm lamp calibration following the
Weekly
maintenance.
Step
Action
, continued
(2 of 2)
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist
ground strap when performing this procedure. For further instructions, refer
to "
" in the beginning of this chapter.
Step
Action
1
Press the STOP button on the instrument to place the system into Stopped
state.
The probes are ready for manual maintenance when system status is
Stopped.
2
Wipe the outside and bottom of any exposed portions of the sample and
reagent probes, and all of the cuvette wash station probes. Use lintless
tissue moistened with 70% isopropyl alcohol or a 70% isopropyl alcohol
pad.
Note: Use a separate pad for each probe.