MC Calibration
Calibration Reports
Troubleshooting Calibration and Result Errors
UniCel DxC Systems Instructions For Use A13914
Page 13-6
October 2005
below defines the Fields in the order they appear for each chemistry on a
typical MC Calibration Report.
Failed Calibration Log
(ERR)
The fields listed below categorize information about a
failed calibration. Multiple errors are shown here for any
chemistry failure. The errors should correlate with any of
the asterisked (*) fields listed within the calibration report.
This section of the report appears only when a calibration
fails.
• Reagent – Identifies the failed chemistry.
• Level – Identifies the calibrator level which failed.
• Condition – In the event that an error occurs, an asterisk
(*) appears next to the value which is out of
specification. The corresponding error flag will print in
this field. Only those errors that FAIL the calibration
will print.
Table 13.5
Cup Module Calibration Report Field Definitions
Field
Description
Rate
Represents the rate of change in ADCs during the reaction.
Rate is used for back-to-back, span and range checks.
Initial Read
Represents the conductance measured by the electrode or
absorbance measured by detector with only reagent present
in the cup.
Final Read
Represents the conductance measured by the electrode or
absorbance measured by detector after the reaction has
taken place.
Set Point
The value assigned to the calibrator level.
Set Point Units
The units assigned to the set point value.
Replicates Used
The field indicating which two of the four calibrator
replicates are used for a particular chemistry.
Span
The difference between the average of the values reported
for rate for each calibrator level. This value is compared to
an expected value; if the calculated span is less than the
expected span, the calibration is flagged.
Table 13.4
ISE Module Calibration Report Field Definitions
, continued
Field
Description
(2 of 2)