Three-Month Maintenance
Replace Syringe Plungers
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-48
October 2005
4
CAUTION
When installing the syringes, do not mix the two different syringe
sizes. The MC and CC sample syringes (100 µL) are located on the
back wall of the instrument. The CC reagent syringe (500 µL) is
located on the right side of the instrument.
To install the assembled syringe:
• Fill the syringe assembly completely full with deionized water.
• Tap the syringe to release bubbles.
• Screw the rounded plunger end onto the adaptor.
• Carefully pull the syringe barrel upward until the syringe Luer-lock
fitting is engaged.
• Turn the syringe barrel and lock in place.
5
Select <OK> when all components have been replaced in the syringe
assemblies.
NOTICE
Always select
<OK>
to return the syringe plunger rods to the Home
position (top) of the syringe unit. Selecting
<Cancel>
closes the pop-up
window without homing the syringes. If
<Cancel>
is selected from the
pop-up window and the Syringe Rod Replacement screen is exited, a
motion error could occur.
6
Select <Close> to return to the Maintenance menu.
Step
Action
, continued
(2 of 2)