Two-Month Maintenance
Change/Clean Air Filters
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-43
10
3
Lower the clean, dry filters back into position on each door, placing the
small filters (1) on top and the large filter (2) on the bottom.
Note: Slide each filter into position so that the metal grating faces towards
the inside of the instrument when doors are closed.
Figure 10.25
Cleaning Filters
4
CAUTION
Do not place a damp filter back on the system. Residual moisture may
cause damage to the system. A second set of filters are supplied with
the system to use an alternate.
Vacuum or brush each filter thoroughly to remove dust. If excessively
dirty, rinse with deionized water. Vigorously shake residual water from
filters. Allow the filter to air dry completely.
Store the clean filters to be reinstalled at the next air filter maintenance.
5
Close all compartment doors.
Step
Action
, continued
(2 of 2)