CTS Wick Replacement Procedure
Install the Wick Clip Assembly and Blade Assembly
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-129
10
Install the Wick Clip Assembly and Blade Assembly
Install the Wick Clip Assembly and Blade Assembly as follows:
Step
Action
1
Install Wick Clip Assembly (1) onto Wash Tower (2). Make sure both ends
of the Wick Clip are secured. Refer to
.
2
Carefully pick up the Blade Assembly and orient end points of Blade with
Slots (3) on Wick Clip. Slowly push the Blade Assembly straight down so
that the Blade goes through the Wick, through the Wash Tower, and
through the alignment slot at the bottom. Refer to
CAUTION
The points on the end of the Blade are very sharp and have been
exposed to potentially biohazardous fluids. To prevent possible injury
or exposure, avoid the Blade area.
3
Lift up the Blade Assembly again as shown in
Assembly back on the plastic tray.
4
Remove the Wick Clip Assembly again as shown in
Examine the bottom of Wick area for any loose fibers. Remove any loose
fibers with scissors.
Note: Loose fibers will clog the drain.
5
Repeat Steps 1 and 2 above.
6
Align Blade Holder (2) to screw holes. Install and tighten the two Blade
Holder Screws (1) with a screwdriver. Refer to
.
7
Install the Cap Piercer Cover (2). Tighten the two cover Screws (1) with a
screwdriver. Refer to
8
Once the Cap Piercer Cover has been installed, select <Cancel> on the
screen to retain the blade count since the Blade was not replaced. If OK is
selected, the system resets the blade count to zero.
9
Select Exit [F10] to exit the Maintenance screen.
10
Place four sample tubes that are capped and empty into an appropriate
sample rack.
• If the system is a DxC 600/800, go to Step 11.
• If the system is a DxC 600i, go to Step 12.