Operational Conditions
Water Requirements
System Description
UniCel DxC Systems Instructions For Use A13914
Page 2-4
October 2005
Water Requirements
Other System Specifications for IEC-1010 Compliance
Table 2.5
Water Requirements
Item
Specification
Flow Rate
0.6 L/min peak flow rate
16 L/hr, minimum continuous flow rate
Temperature
+15°C to +25°C
Water quality
NCCLS Type II deionized water, except for total
bacteria count < 10 cfu/mL
Water pressure
Deionized water entering the system must be 30–90
psig.
Table 2.6
IEC-1010 Specifications
Item
Specification
Pollution Degree
2
EN55011
Meets Class A
Maximum Sound Pressure
≤
65 dBA average over 8 hours with covers down at 1
meter away from the instrument at 25°C
Maximum Leakage Current DxC 600: 222 µA at 240V, 50Hz
DxC 800: 240 µA at 240V, 50Hz