Calibrator Assignment
UniCel DxC Systems Instructions For Use A13914
Reagent Load/Calibration
October 2005
Page 5-15
5
Calibrator Assignment
To calibrate a reagent, first assign the rack/position or calibrator ID. The assignments
can be reused or new assignments can be programmed with each new calibration.
Note: A reserved rack must be used when placing a calibrator cup in a tube.
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select Assign [F7].
3
Use the down arrow option button to open list of calibrators.
4
Select the calibrator.
5
Type in the calibrator’s ID or rack/position number.
• If using a bar code label, the label must match the assigned calibrator ID.
• If a rack/position is assigned, the calibrator must be placed in this
assigned position for all calibrations.
6
Repeat Steps 3 and 4 to assign additional calibrators
7
When calibration assignment is complete, select <CLOSE>.