Summary of Precautions
Sample Bar Code Reader Precautions
UniCel DxC Systems Instructions For Use A13914
General Information
October 2005
Page 1-11
1
Sample Bar Code Reader Precautions
Do not tamper with or remove the housing of the Sample Bar Code Reader.
Sample Syringe Replacement Precautions
When installing the syringes, do not mix the two different syringe sizes. The MC and
CC sample syringes (100 µL) are located on the back wall of the instrument. The CC
reagent syringe (500 µL) is located on the right side of the instrument.
Samples NOT received by Host Precautions
Do not clear samples until results are received at the host and printed. Clearing
samples manually or through host programming, may cause results to be received at
the host and printed at the DxC without the sample ID. Depending on the host
implementation, this can cause lost sample results or sample results which merge with
other sample results producing duplicate tests or added tests.
Samples Use Precautions
Do not use the same sample run on an DxC system for analysis of analytes for which a
small quantity of carryover could greatly increase the results (for example, TBhCG).
If your system has a Closed Tube Sampling (CTS) option, Do not run samples in cups
on top of a sample tube. This would give motion errors. As an alternative, put the
sample in a nesting cup on a tube in a reserved rack.
System Backup Precautions
After inserting a diskette into the floppy disk drive, selecting <OK> when prompted
will prepare the floppy disk by ERASING it before copying data. Be sure the diskette
does not contain critical data that is not available from another source.
System Configuration Change Precautions
Changes to the System Configuration Data should only be done at the request or at the
direction of Beckman Coulter, Inc. Entry of incorrect information will lead to system
errors.
System Restore Precautions
System Parameter and Alignment data can be restored from the backup floppy disks
onto the system, however performing the Restore function deletes some or all files
(depending on the areas restored) from the hard drive.
Urine Sample Precautions
After analysis of ten consecutive urine electrolytes, run one replicate of electrolytes on
Synchron Calibrator Level 2 in the serum mode. This will minimize the potential for
chloride drift due to matrix effects of urine samples.