Prime
Introduction
Utilities
UniCel DxC Systems Instructions For Use A13914
Page 12-2
October 2005
Prime
Introduction
The Prime feature of the Utilities menu permits the user to manually request a prime.
An entire subsystem can be primed. A prime can also be requested on an individual
component (refer to the tables below). This feature is only available when the system
status is in Standby.
Table 12.1
A Prime for the CC Subsystems
CC Subsystems
Function
Reagent Delivery Subsystem
Washes the reagent probes and mixer, primes the
reagent syringe.
Sample Delivery Subsystem
Washes the CC sample probe and mixer, primes
the sample syringe.
Cuvette Wash Station
Primes cuvette wash Probes 1 and 2 with wash
and Probe 3 with water.
Table 12.2
A Prime for the Hydropneumatics Subsystem
Hydropneumatics Subsystem
Function
Fill canisters and reservoirs
Fills water, wash, and diluted wash.
CC Drain waste sump
Drains waste sump and exit sump.
CC Drain Waste B sump
Drains Waste B sump and exit sump.
Drain gravity drain sump
Drains the gravity sump.
Table 12.3
A Prime for the Cups of the MC Subsystems
MC Subsystems
Function
Cups
ALBm (DxC 800 only)
Primes Albumin Reagent and/or water.
CREm (DxC 800 only)
Primes Creatinine Reagent and/or water.
GLUCm (DxC 600 and DxC
800)
Primes Glucose Reagent and/or water.
PHOSm (DxC 800 only)
Primes Phosphorus Reagent and/or water.
TPm (DxC 800 only)
Primes Total Protein Reagent and/or water.
BUNm/UREAm (DxC 800
only)
Primes BUN/UREA Reagent and/or water.