As-Needed/As-Required Maintenance
EIC Port Cleaning
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-86
October 2005
EIC Port Cleaning
Flushing of the EIC ports should be performed if the Electrolyte Injection Cup (EIC)
becomes plugged and overflows, or as part of troubleshooting for erratic electrolyte
results. The following items should be available before starting this procedure:
• 10% bleach solution (1 part sodium hypochlorite 5.25% combined with 9 parts
deionized water). Prepare fresh. Use within 24 hours.
• 70% isopropyl alcohol
• 20 or 30 mL irrigation syringe
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist
ground strap when performing this procedure. For further instructions, refer
to "
" in the beginning of this chapter.
Step
Action
1
Loosen the two Phillips screws on the ISE module cover and remove the
cover. Release the locking pin on the ISE module and raise the ISE module
to a locked position.
2
Prepare ISE for maintenance and disable autoprime by selecting:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <3> ISE Service.
Select <Continue> to prepare for maintenance.
(1 of 7)