Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-50
October 2005
0x04B2001E
4
MC Sample Probe
Obstruction
Sample ID: sample
id, Rack ID: rack,
Position: position.
This occurs on first
attempt
CAU: MC Sample
Probe Obstruction
Message ID:
0x04B2001E
Sample ID: sample id
Rack: rack
Position: position
Description: This error occurs
when the MC sample probe ODC
(Obstruction Detection &
Correction) sensor indicates that the
probe is obstructed during sample
aspiration.
• No operator action is necessary.
• The system will attempt a second
aspiration from this sample.
0x04B2001F
4
Unrecovered MC
Sample Probe
Obstruction
CAU: Unrecovered MC
Sample Probe
Obstruction
Message ID:
0x04B2001F
Description: This error occurs
when there was an Obstruction
present at the MC sample probe.
The probe tried to clear the
obstruction by flushing, but the
obstruction could not be removed.
The MC side of the system goes to
a "Stopped" state.
• Flush the MC Sample probe as per the
"As Needed" Probe Flushing
Maintenance Procedure.
• If flushing probe is not successful,
replace MC Sample probe.
• If probe is clear and error continues,
check for pinched Sample line tubing.
• If the error continues call Beckman
Customer Support.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action