As-Needed/As-Required Maintenance
Replacing the Chloride Electrode Tip
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-71
10
3
Remove the retainer nut from the old electrode. Remove the quad-ring (1)
from the tip of the old electrode. Inspect the electrode port if the quad-ring
is not on the electrode. Discard the quad-ring.
Figure 10.45
4
Unscrew the old tip from the electrode assembly. Discard the O-ring and
the old electrode tip.
5
Install a new O-ring on chloride electrode tip and screw the tip onto
electrode body.
6
Install the retainer nut and a new quad-ring on the electrode and insert the
new electrode into electrode port. Turn electrode retainer nut clockwise
until finger-tight.
7
Gently pull on electrode body to test for proper seating of electrode. The
electrode assembly should not move. If electrode moves, remove it and
install it again. Check for an extra or missing quad-ring if installation is
difficult.
8
Reposition flow cell on the mounting panel and tighten the one retaining
screw. Reconnect electrode cable to the appropriate terminal being careful
not to bend the connecting pin.
9
Select <OK> in the Maintenance confirmation pop-up window when the
electrode service has been completed and the electrode has been properly
replaced.
NOTICE
Reagents will immediately prime through the flow cell.
While the system is priming, observe the low cell for leaks. Stop the
priming if any leaks are noticed and correct the problem.
Step
Action
, continued
(2 of 3)