How to Use Reserved Racks
Processing Unreadable Bar Coded Samples
Preparing Samples for Analysis
UniCel DxC Systems Instructions For Use A13914
Page 3-12
October 2005
Processing Unreadable Bar Coded Samples
If a sample bar code is unreadable and the LIS needs to be queried, use manual
assignment. A manual assignment is rack and position and a Sample ID with no other
information. The DxC will query the LIS for tests and demographics.
Note: If previous programming appears, it may be necessary to clear it.
Processing Samples Manually
Manual programming is used when there is no LIS or the LIS is not available.
Samples may or may not be bar coded. The sample is identified by a Sample ID and/
or rack and position. The user selects tests for programming.
The following items require manual programming:
• samples without bar codes
• samples with bar codes and no sample programming
If...
Then...
Individual Samples,
• Select the Samples icon from the menu bar,
• Type rack and position,
• Type Sample ID,
• Save [F10].
Several samples in the same rack, • Select the Samples icon from the menu bar,
• Select Rack [F8],
• Select the rack,
• Type Sample ID(s),
• Select <OK>.
Step
Action
1
Select the Samples icon from the menu bar.
2
Before programming, check for previous programming.
• Select Racks [F8].
• Select the rack number and select <OK>.
• If the rack has programming that is no longer required, select Clear
[F1].
• Type the Sample IDs into the Sample ID(s) field.
• Select <OK>. A second screen will open. Select <OK> again to clear
the data.
3
Select Program [F10].
(1 of 2)