Non-Linear and Multipoint Calibrations
Calibration Report
Troubleshooting Calibration and Result Errors
UniCel DxC Systems Instructions For Use A13914
Page 13-14
October 2005
Calibration Report
The CC Non-Linear and Multipoint Calibration Reports can be a useful tool for
identifying issues.
below provides definitions of specific calibration fields
found on those two reports.
Table 13.7
Non-Linear Single Point and Multipoint Calibration Reports
Field Definitions
Field
Description
Reagent Identifiers
The following fields identify specific cartridge
information.
• Rgt – Indicates which chemistry is being calibrated.
• Units – Shows selected units for reference.
• S/N – Specifies the serial number of the reagent
cartridge being calibrated.
• Lot – Identifies the lot number of the specific cartridge.
• Loc – Identifies the location of the reagent cartridge on
the reagent carousel.
• Math Model – Identifies the specific Math Model used
to calculate the curve parameters.
Cuv
The cuvette number on the reaction carousel in which the
reaction took place.
Level
The level of calibrator being run.
Set Points
The target concentration value for the calibrator level.
Blank
The reagent blank absorbance when it is an ENDPOINT
chemistry or the reagent blank rate (absorbance per unit
time) when it is a RATE chemistry.
Reaction
The reaction absorbance when it is an ENDPOINT
chemistry or the reaction rate (absorbance per unit time)
when it is a RATE chemistry. For ENDPOINT 2 or RATE
2 reactions, the blank absorbance must be subtracted from
this value to obtain the net delta absorbance.
Recovery
The concentration obtained when the final calibrator
absorbencies are applied to the calculated curve.
Deviation
The difference between the obtained absorbance value and
the curve in terms of absorbance.
Standard Deviation of
the Curve
The standard deviation of all the calculated deviation
values. An indication of curve fit.
(1 of 3)