![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768033.webp)
Sample Handling System
Introduction
UniCel DxC Systems Instructions For Use A13914
System Description
October 2005
Page 2-7
2
Sample Handling System
Introduction
The Sample Handling system is composed of the following components:
• Sample racks
• Autoloader/Offload track
• Priority load position
• Shuttle
• Bar code reader
• Cap piercer assembly (optional)
• Sample Carousel
• Sample probe/mixer assemblies
• Wash cup assemblies
• Obstruction detection assemblies
The sample handling module is used to load samples onto the system, provide samples
for analysis, and provide temporary storage of completed samples. A detailed
description of each component is presented in the following paragraphs.
Sample Racks
A sample rack is a high-strength, plastic, centrifugable holder designed to house up to
four samples. There are four sizes of racks with each size capable of holding one
length/width combination of primary sample tubes in addition to accepting sample
cups. (Refer to
Sample racks accept the following tube and cup sizes:
NOTICE
Adapters are provided to adapt various sized sample tubes (secondary tubes)
to the short racks. These adaptors must only be used in racks designated as
reserved. The reserved rack feature is described in
Table 2.7
Sample Racks
Rack
Accepts These Cups and Tubes
13 × 75
12 × 75 mm tubes
13 × 75 mm tubes
0.5 mL cups
2.0 mL cups
(1 of 2)